Promega Corporation

Measuring the QuantiFluor™ dsDNA System Using the QuantiFluor™-ST Fluorometer

  • Share
  • Print
  • Email
  • Download PDF
  • Article
  • Comments & Ratings

Abstract

Traditional spectrophotometric assays cannot determine DNA concentration below 2µg/ml; however, many isolated DNA samples have concentrations well below that level. The QuantiFluor™ dsDNA System provides a fast, easy and sensitive method for determining DNA concentrations as low as 2.5ng/ml (or 0.5ng/well). The QuantiFluor™ dsDNA System contains a fluorescent DNA-binding dye that enables sensitive and specific quantitation of small amounts of double-stranded DNA (dsDNA) in solution. The dye shows minimal binding to single-stranded DNA (ssDNA) and RNA, allowing specific quantitation of dsDNA.

This Application Note describes the protocol for using the QuantiFluor™ dsDNA System with the QuantiFluor™-ST Fluorometer and the PCR Tube Adapter. The QuantiFluor™ PCR Tube Adapter allows for sample volumes as little as 100µl without sacrificing instrument sensitivity.

Publication Date: 11/12

Download Article (200 KB)

  • Warning
  • It appears that you have Javascript disabled. Our website requires Javascript to function correctly. Please enable Javascript to experience our website.

Scientists at Your Service

We offer a range of services to help you succeed using Promega technologies. From product training to set up of automated systems and development of custom applications—our scientific support goes beyond the basics.

Ask us! We are here to help you.

Click Here for Technical Questions and Ordering Support »